Flow in Liters per Minute (LPM) -> 1 LPM -> 1000 ml/min (1000 ml/min)
Ozone Concentration in Gamma -> 1 gamma -> 1µg/ml -> 1 mg/l -> 1gr/m3

Temp: 21 C
Feed gas: Oxygen/Med
Pressure: 100.1 kPa
Note: YY/5.5/120/714/BT/15C/12VDC
Date: Aug 22, 1999
Model: GE30/12VDC
Serial #: 990822.a
File:GE30DC01.XLS
Flow in Liters Per Minute (LPM).  Ozone concentration in GAMMA ( 1 µg/ml = 1 mg/l )

FLOW FLOW
------- OZONE CONCENTRATION REGULATOR SETTING ----------
LPM cc/min 1 2 3 4 5 6 7 8 9 10

1/32 31 20 26 33 40 46 52 54 56 63 70
1/16 62 9 12 16 20 24 28 32 35 40 44
1/8 125 5 7 10 13 16 19 19 25 28 31
1/4 250 3 4 6 7 9 10 5 14 16 17
1/2 500 1 2 3 3 4 5 6 7 8 9
3/4 750 1 1 2 2 3 3 4 5 6 6
1 1000 1 1 2 2 2 2 3 3 4 4

Test - concentration regulator setting #1, #2, #4, #6, #8 and #10 with oxygen flow from 1/32 LPM to 1 LPM (bold print).
Estimate - setting #3, #5, #7, and #9 (plain print)

Please see out page on How to Read the Ozone Test Report for an explanation of the above report.